With the aging of the population, the rapid increase in the incidence of osteoporosis, and increased interest in healthy life, DEXA tests are increasing every year. This is the use of medical radiation, which plays an important role in early detection of osteoporosis and is a cause of medical exposure. Accordingly, using self-made shields, the radiation dose rate before and after shielding against other areas except the inspection area was measured and the reduction rate was calculated. As a result, the reduction rate before and after shielding was found from the lowest 51.61% to the highest 97.60%. This can minimize exposure to areas unrelated to the purpose of measuring by using effective shields during bone mineral density examination, and also maintain the accuracy of examination values.
한국어
인구의 고령화와 더불어 골다공증 발생률이 급증하고 건강한 삶에 대한 관심이 높아지면서 DEXA 검사가 매년 증가하 고 있다. 이는 의료용 방사선을 사용하는 것으로 골다공증을 조기발견에 중요한 역할을 하는 동시에 의료피폭의 원인이 되고 있다. 이에 자체 제작한 차폐체를 이용하여 검사부위를 제외한 타 부 위에 대한 차폐 전ㆍ후의 방사선량율을 측정하 고 감소율을 계산하였다. 그 결과, 차폐 전ㆍ후의 감소율은 최저 51.61%에서 최고 97.60%로 나타났다. 이는 골밀도 검사 시 효과적인 차폐체의 사용으로 촬영목적과 관계가 없는 부위의 피폭을 최소화 할 수 있고, 검사 수치의 정확도를 유지할 수 있을 것이다.
목차
Abstract Ⅰ. 서론 Ⅱ. 실험 및 방법 1. DEXA 장비 및 선량계 2. 환자 차폐체의 설계 및 제작 3. 측정 조건 및 방법 4. 방사선량율의 계산 Ⅲ. 결과 1. 차폐체 사용 전·후 방사선량율 2. 부위별 차폐체 사용 전·후 감소율 Ⅴ. 참고문헌 요약