We checked the existence of bacteria at DR equipment, radiation generator and cassette in ‘D university’ to propose optimized management method for bacteria and disinfection at DR equipment, radiation generator and cassette at used in radiographic laboratory of the department of radiology. The infected region is treated by alcohol and cresol and compared before spraying and after spraying for 1, 3, 5 and 10 minutes. As time goes on, amount of bacteria colony decreases significantly. The Cresol is useful to disinfect the diminutive devices. However alcohol disinfection, at least 5 minutes later, and hand washing before practical training and shooting image is recommended due to the big and fixed radiography equipment which cause the soakage.