Purpose : Fixed way of mAs previously Low Extremity Computed Tomography Angiography(LECTA) examination were used. Automatic Current Selection(ACS) to use for the purpose of reducing the dose when Low Extremity Computed Tomography Angiography examining patients. Materials and methods : Were analyzed from July 2011 to July 2012 MDCT examination of Dose Length Product(DLP) LECTA 116 Case. It was defined as previous inspection methods(Old protocol). CT workstation is set to 100 mAs and 150 mAs protocol based on the patient's weight 70kg examined by LECTA. We defined as 'New protocol' that applies to ACS. The data collection period are 76 cases from October 2012 to January 2013 Results : 1. Average Total DLP of ’ Old protocol’ is 3602.943 mGy*cm. 2. Average Total DLP of ’ New protocol’ is 1762.977 mGy*cm. 3. Due to the 'New Protocol' use of Total DLP was reduced by approximately 51 %. Phase-specific dose reduction is as follows. Pre(33.62 %), Artery(64.63 %), Delay(49.0 %) 4. Using One way ANOVA Analysis of fluctuations obtained DLP is as follows. ‘Old protocol’, ’New protocol’ a value of P <0.001, P = 0.882 values were obtained. Conclusions : Dose reduction of 51 % is a useful study that proves. The results obtained using the ACS, the effects of a dose reduction of 51 % was obtained. Therefore, it has been proven to be a useful way. Statistics using SPSS version came out of the 'Old protocol' P-value P <0.0001. This result means that the DLP a large difference values. On the other hand, The results of the 'New protocol' was P = 0.882. These results means to that small and regularly was fluctuations of the dose. The use of ACS, you can get a reduction of the dose and will able to get the effect of reducing the dose errors.

목차

Abstract
Ⅰ. 서론
Ⅱ. 대상 및 방법
1. 영상의 획득
2. 자료의 수집
3. 선량의 비교
Ⅲ. 결과
1. 기존 검사방법(Old protocol)
2. 새로운 검사방법(New protocol)
3. 통계
Ⅳ. 고찰 및 결론
1. 고찰
2. 결론
참고문헌

키워드

Low Extremity CT Angiography Automatic Current Selection Dose Length Product

저자

  • 안형택 [ Hyeong-Theck An | 강동경희대학교병원 영상의학과 ]
  • 김재열 [ Jae-Yeol Kim | 강동경희대학교병원 영상의학과 ]
  • 여운식 [ Woon-Sik Yeo | 강동경희대학교병원 영상의학과 ]
  • 박용성 [ Yong-Sung Park | 강동경희대학교병원 영상의학과 ]
  • 이귀원 [ Gui-Won Lee | 동남보건대학교 방사선학과 ]
  • 이종웅 [ Jong-Woong Lee | 강동경희대학교병원 영상의학과 ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 15 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장