Purpose : Automatic exposure system(AOP mode) in DR Mammography of the STD and DOSE apply to women of childbearing age, the average glandular dose(AGD) and image quality by comparing was to demonstrate the usefulness of DOSE. Materials and Methods : Of the under 40 age 108 patients who visit to our hospital and examined STD and DOSE mammography from January 2008 to July 2013. AGD was obtained by DICOM header information provided by GE Senographe DS. STD and DOSE images were evaluated with obtained patients, Image J program was compared by calculating the SNR. Results : The average AGD of DOSE mode was 0.99mGy, and which decreased by 19% comparing to that of STD mode which was 1.18mGy. The two the average AGDs indicated statistically significant difference(p<.01). The average SNR of STD was 40.26, DOSE displayed, and to 39.68 in, there was no statistical significance. Results : The average AGD using DOSE mode which is one of DR mammography AOP modes decreased by comparing to that of STD mode showing no difference in image quality. The use of DOSE is considered to be useful.