방사선(학)과 Apron 성능평가 및 관리 조사 - 대구, 경북지역 4개 대학교 -
The study of evaluation about the apron quality management and shield efficiency in 4 university laboratories of dae-gu, Kyungpook area
This study evaluated the apron quality management and shield efficiency in the education laboratory for radiology students. For making a proposal of effective control of lead apron, Eighteen aprons, used in 4 university laboratories of dae-gu, Kyungpook area, were analyzed regarding their uniformities, transmission doses, lead equivalent and shield efficiencies. In conclusion, the lead apron uniformity was generally good. Also, the shield efficiency corresponding to lead equivalent was high. However, the quality control of lead apron was not satisfactory. Since there is a potentially possible chance for students to be exposed to a radiation during practical radiation-training in the education laboratory, the quality-control maintenance of the lead apron should be unceasingly maintained for the radiation protection.
목차
Abstract Ⅰ. 서론 Ⅱ. 실험 1. 실험장치 및 기기 2. 실험방법 Ⅲ. 결과 Ⅳ. 결론 참고문헌
키워드
Apron납당량(mmPb)차폐율(%)
저자
최범희 [ Beom-hee Choi | 대구보건대학교 방사선과 ]
김용국 [ Yong-gug Kim | 대구보건대학교 방사선과 ]
박서현 [ Seo-hyun Park | 대구보건대학교 방사선과 ]
신용일 [ Yong-il Shin | 대구보건대학교 방사선과 ]
유영준 [ Young-jun You | 대구보건대학교 방사선과 ]
김예리 [ Ye-ri Kim | 대구보건대학교 방사선과 ]
박창희 [ Chang-hee Park | 대구보건대학교 방사선과 ]
Corresponding Author