To be investigated and measures to reduce the medical exposure of patients to change the Epidural Injection time Fluoroscopy Pulse Rate in this study. Was conducted in 50 patients who underwent surgery Epidural Injection performed by interventional care of Konkuk University Hospital from January to April 2013. The treatment time with the change of Pulse rate, is measured in minutes fluoroscopy time, and measured the area dose (μGy㎡) and depth dose (mGy). Using the Image J program, to measure the PSNR and SNR. The fluoroscopy time as a result surgery, there was no significance in the statistical analysis, and depth dose is 34.3 to 34.9%, was reduced from 35.8 to 38.7% the area dose. It is possible to reduce the appropriate Pulse rate, to reduce the dose without statistical analysis significance fluoroscopy time.
목차
Abstract Ⅰ. 서론 Ⅱ. 대상 및 방법 1. 대상 2. 방법 Ⅲ. 결과 1. 시술시간 2. Fluoroscopy rate 변화에 따른 PSNR측정 3. 심부선량과 면적선량 5. Pulse rate별 선량 감소율 Ⅳ. 고찰 Ⅴ. 결론 참고문헌
키워드
Pulse ratePSNRDAPdose