The Relationship of Radiation Dose and Image Quality According to the Condition of Chest PA

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 13 Number 4 (2011.12)바로가기
  • 페이지
    pp.165-169
  • 저자
    Jin Hyun Son, Jung Whan Min, Byung Sam Kang, Kyung Rae Dong
  • 언어
    영어(ENG)
  • URL
    https://www.earticle.net/Article/A483023

※ 기관로그인 시 무료 이용이 가능합니다.
※ 학술발표대회집, 워크숍 자료집 중 4페이지 이내 논문은 '요약'만 제공되는 경우가 있으니, 구매 전에 간행물명, 페이지 수 확인 부탁 드립니다.

4,000원

원문정보

초록

영어
The purpose of this study is to compare the measurement result of radiation dose by using standard thoracic phantom and ionization chamber to advice proposal in the shooting condition of chest PA projection at hospitals recently. And to understand the change between radiation dose and resolution in different conditions. The period this study was from August 2010 to September 2010 and the subjects of the study was 3 general hospitals, 4 personal hospitals and 1 laboratory at the college. Finally we study with 6 DR, 1 CR, and 4 F/S equipments. Most hospitals met advice proposal, but some of the hospitals exceed advice dose from the result of our study. We can lower radiation dose about 25% when kVp is lowered about 20% in DR equipment. And we can lower radiation dose about 50% when mAs is lowered about 35%. The image quality was similar to the original in the study. Most hospitals which exceed advice dose were personal hospitals. The reason why it happened is that radiation dose for chest PA projection at personal hospitals is higher than general hospitals and the personal hospitals´ equipments are older than general hospitals´ equipments. We guess that patients´ radiation dose of chest PA projection can be lowered from the result.

목차

Abstract
Ⅰ. Introduction
Ⅱ. Material & Method
1. Object and examination equipment
2. Examination method
Ⅲ. Result
Ⅳ. Discussion
Ⅴ. Conclusion
References

키워드

Standard thoracic phantom Ionization chamber Radiation dose chest PA

저자

  • Jin Hyun Son [ Department of Radiological Technology, Shingu University ]
  • Jung Whan Min [ Department of Radiological Technology, Shingu University ]
  • Byung Sam Kang [ Department of Radiological Technology, Shingu University ]
  • Kyung Rae Dong [ Department of Radiological Technology, Gwangju Health College University/Department of Nuclear Engineering, Chosun University ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 13 Number 4

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장