※ 기관로그인 시 무료 이용이 가능합니다.
※ 학술발표대회집, 워크숍 자료집 중 4페이지 이내 논문은 '요약'만 제공되는 경우가 있으니, 구매 전에 간행물명, 페이지 수 확인 부탁 드립니다.
4,000원
원문정보
초록
영어
Dynamic range on the digital detector can be a representation to the ratio of maximum and minimum of pixel value. Wide dynamic range and post processing ability of the digital detector made difficult to recognize visually to high or low dose images. We were evaluated a change of mean pixel value on the original and processed image, when we controlled the kVp, mA, exposure time on the digital detector. On the kVp of a constant condition, we were acquired an original and processed image by changes of mA, exposure time. According to the thickness of the subject under the same conditions, to determine a relation of pixel value and X-ray intensity, we used an aluminum step wedge. When mA and exposure times were changed under the kVp of a constant condition, the X-ray intensity was decreased by the reduction of the mean pixel value. In addition when kVp was increased in a constant condition of mAs, the mean pixel value was increased according to the increment of the X-ray intensity. Therefore, low kVp, high mA and short exposure time were a way to reduce a patient dose.