Yeong-Hyeon Park, Chan-Young Park, Yu-Seop Kim, Hye-Jeong Song, Jong-Dae Kim
언어
영어(ENG)
URL
https://www.earticle.net/Article/A251420
※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.
원문정보
초록
영어
This paper presents the analysis of the steady state error and the time delay of the temperature of the chip sensor in a micro PCR chip and that of the reagent. The chambers with the various heights and sizes are compared in order to examine the differences in temperature and speed. A NTC-thermistor was employed as the chip sensor and the thermocouples were used to examine the temperature of the reagent in the chamber. From the result, we confirmed that the steady state error was negligible for all of constructions and the delay times showed no difference in the range of 60–72°C, while in the other ranges, the speed was higher for the lower height construction.
목차
Abstract 1. Introduction 2. Materials and Methods 2.1. Micro PCR Chip 2.2. Installation of the Thermocouples 2.3. Temperature Measurement Experiment 3. Results 4. Conclusion Acknowledgments References