Bag-Valve-Mask의 사용방법에 따른 일회호흡량과 평균기도압의 변화 연구
Differentiation of tidal volume & mean airway pressure with different Bag-Valve-Mask compression depth and compression rate
Purpose : The purpose of this study is to get basal user guidelines of safer bag-valve-mask application on patient with normal pulmonary patho-physiologic condition. Methods : This study was accomplished by pre-qualified 25 EMS junior grade students. Participants were instructed randomly compress bag to one-third, half and total and also with differesnt compression speed. Resultant tidal volumes and mean airway pressures obtained in RespiTrainer were analysed in relation to the each compression depth and rate. Results : Demographic difference does not affect tidal volume with any compression depth and rate change. Increasing compression depth is correlated with tidal volume increasement at any compression rate and also with mean airway pressure. If the compression depth is same, compression rate change did not affect significantly the resultant tidal volume or mean airway pressure. Conclusion : Hand size, Experience, BMI dose not affect tidal volume. Compress the 1600 ml bag half to total amount is safe way to offer sufficient tidal volume without risky high airway pressure delivery to patient airway who with normal lung patho-physiologic condition.
목차
Abstract I. 서론 II. 연구방법 1. 연구대상 2. 연구방법 III. 연구결과 IV. 고찰 V. 결론 참고문헌
한국응급구조학회 [The Korean Society of Emergency Medical Technology]
설립연도
1997
분야
의약학>응급의학
소개
복합 구조적인 산업사회의 발전에 따라 안전사고와 응급상황에서의 구조구급요청자들이 증가하고 있으나 우리나라의 응급의료체계는 아직 그 수준을 따라가지 못하고 있는 실정이다. 이에 우리나라의 응급의료체계를 확립하고 구조구급에 관한 기초학문의 발전에 기여하고자 학회를 창립한다.
간행물
간행물명
한국응급구조학회지(구 한국응급구조학회논문지) [The Korean Journal of Emergency Medical Services ]