Earticle

현재 위치 Home

논문

TiN 코팅된 Ti 및 Ti-6Al-4V합금의 부식거동
Corrosion behaviors of Cp-Ti and Ti-6Al-4V alloys by TiN coating

첫 페이지 보기
  • 발행기관
    대한치과기공학회 바로가기
  • 간행물
    대한치과기공학회지 바로가기
  • 통권
    Vol.30 No.1 (2008.07)바로가기
  • 페이지
    pp.25-31
  • 저자
    이순현, 정용훈, 최한철, 고영무
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A89897

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
Cp-Ti and Ti-6Al-4V alloys commonly used dental implant materials, particularly for orthopaedic and osteosynthesis because of its suitable mechanical properties and excellent biocompatibility. This alloys have excellent corrosion behavior in the clinical environment. The first factor to decide the success of dental implantation is sufficient osseointegration and high corrosion resistance between on implant fixture and its surrounding bone tissue. In this study, in order to increase corrosion resistance and biocompatibility of Cp-Ti and Ti-6Al-4V alloy
that surface of manufactured alloy was coated with TiN by RF-magnetron sputtering method. The electrochemical behavior of TiN coated Cp-Ti and Ti-6Al-4V alloy were investigated using potentiodynamic (EG&G Co, PARSTAT 2273. USA) and potentiostatic test (250mV) in 0.9% NaCl solution at 36.5 1 .
These results are as follows :
1. From the microstructure analysis, Cp-Ti showed the acicular structure of -phase and Ti-6Al-4V showed the micro-acicular structure of + phase.
2. From the potentiodynamic test, Ecorr value of Cp-Ti and Ti-6Al-4V alloys showed -702.48mV and -319.87mV, respectively. Ti-6Al-4V alloy value was higher than Cp-Ti alloy.
3. From the analysis of TiN and coated layer, TIN coated surface showed columnar structure with 800 nm thickness.
4. The corrosion resistance of TiN coated Cp-Ti and Ti-6Al-4V alloys were higher than those of the non-coated Ti alloys in 0.9% NaCl solution from potentiodynamic test, indicating better protective effect.
5. The passivation current density of TiN coated Cp-Ti and Ti-6Al-4V alloys were smaller than that of the noncoated implant fixture in 0.9% NaCl solution, indicating the good protective effect resulting from more compact and homogeneous layer formation.

목차

Abstract
 Ⅰ. 서론
 Ⅱ. 연구 재료 및 방법
  1. 연구 재료 및 미세구조 관찰
  2. TIN코팅
  3. 부식 실험
 Ⅲ. 결과 및 고찰
 Ⅳ. 총괄 및 고안
 Ⅴ. 결론
 참고문헌

키워드

Cp-Ti Ti-6Al-4V alloys TiN coating osseointegration Corrosion.

저자

  • 이순현 [ Soon-Hyun Lee | 대명치과기공소 ]
  • 정용훈 [ Yoong-Hun Jung | 조선대학교 치과대학 치과재료학교실 ]
  • 최한철 [ Han-Chul Choi | 조선대학교 치과대학 치과재료학교실 ]
  • 고영무 [ Yeong-Mu Ko | 조선대학교 치과대학 치과재료학교실 ]

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한치과기공학회 [The Korean Academy Of Dental Technology]
  • 설립연도
    1979
  • 분야
    의약학>치의학
  • 소개
    본 학회는 치과기공에 관한 모든 분야의 연구와 회무를 총괄하기 위하여 세분화된 분과학회의 회원 상호간 학술교류와 정보를 수집하고, 보수교육 및 학술대회 준비, 진행과 학회지 발간등의 업무를 효율적으로 수행함을 목적으로 한다.

간행물

  • 간행물명
    대한치과기공학회지 [THE JOURNAL OF KOREAN ACADEMY OF DENTAL TECHNOLOGY]
  • 간기
    계간
  • pISSN
    1229-3954
  • eISSN
    2288-5218
  • 수록기간
    1979~2026
  • 등재여부
    KCI 등재
  • 십진분류
    KDC 515 DDC 617

이 권호 내 다른 논문 / 대한치과기공학회지 Vol.30 No.1

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장