Assessment of Shielding Effectiveness for Radiation Dose Reduction in Adult Brain Photon-Counting Computed Tomography : A Polymethyl Methacrylate Phantom Study
Sung-Yi Tsai, Dong Shen, Xinmiao Wang, Lianzheng Su, Yuxi Ge
언어
영어(ENG)
URL
https://www.earticle.net/Article/A492164
원문정보
초록
영어
Background: The efficacy of lead-equivalent shielding in modern photon-counting computed tomography (PCCT) for adult brain imaging remains under debate and therefore requires quantitative evaluation under controlled conditions. This study aimed to quantify the radiation dose reduction to out-of-field radiosensitive organs and the effective dose (E) achievable using supplementary lead shielding during adult brain PCCT, employing a polymethylmethacrylate (PMMA) anthropomorphic phantom. Materials and Methods: A PMMA phantom was scanned using a first-generation PCCT system with a standard adult brain protocol under two conditions: unshielded and shielded. Shielding consisted of a 0.5 mm lead-equivalent neck collar and apron, which were applied after scout acquisition. Organ equivalent doses were measured at 55 anatomical sites using thermoluminescent dosimeters. The E was calculated using tissue weighting factors from International Commission on Radiological Protection (ICRP) Publication 103 and was compared with doselength product (DLP)-based estimates. Results and Discussion: Lead shielding resulted in statistically significant dose reductions (all false discovery rate-adjusted q<0.05) in multiple out-of-field organs, including the breast (83.3%), esophagus (88.5%), lungs (85.2%), and thyroid (44.5%). The overall E was reduced by 8.7%, from 2.98 mSv to 2.72 mSv. A substantial methodological discrepancy was observed, as DLPbased E estimates were 47%–52% lower than those derived from phantom measurements using ICRP Publication 103. Conclusion: This phantom study demonstrates that lead-equivalent shielding can significantly reduce radiation dose to out-of-field organs during adult brain PCCT. However, achieving this benefit is critically dependent on proper procedural implementation. Shielding must be applied after scout (localizer) acquisition to avoid interference with the scanner’s automatic exposure control system. The marked discrepancy between dose estimation methods indicates that DLPbased values, while useful for population-level benchmarking, have inherent limitations for precise, protocol-specific optimization. These findings support the targeted and protocol-appropriate use of shielding for vulnerable patients, in accordance with the As Low As Reasonably Achievable (ALARA) principle.
목차
ABSTRACT Introduction Materials and Methods 1. PCCT Units and Imaging Protocol 2. PMMA Phantom 3. Thermoluminescent Dosimeters 4. Radiation Dose Assessment: Organ and Effective Dose Calculations 5. Accuracy and Precision Evaluation of HT Measurements 6. Statistical Analysis Results and Discussion 1. Organ Dose (HT) under Unshielded and Shielded Conditions 2. Effective Dose: Shielding Effects and Methodological Comparisons 3. Comparison with Other Studies 4. Considerations on Shielding Implementation and AEC Interaction 5. Clinical Relevance Conclusion Article Information References
대한방사선방어학회 [Korean Association For Radiation Protection]
설립연도
1975
분야
자연과학>기타자연과학
소개
회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다.
1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최
2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포
3. 방사선방어에 관한 학술의 국제교류 및 협력
4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역
5. 방사선방어에 관한 조사 및 연구용역
6. 회원의 연구활동을 위한 제반협조
7. 기타 본 학회의 목적 달성에 필요한 사항
간행물
간행물명
방사선방어학회지 [Journal of Radiation Protection and Research]