This study was conducted to confirm the antimicrobial activity and persistence of the disinfection cap containing 70% isopropyl alcohol attached to the needleless connector. Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli , Pseudomonas aeruginosa, Acinetobacter baumannii , Candida albicans , Candida glabrata were allowed to stand at room temperature for 1min, 2hrs, 4hrs, and 7 days after inoculation of 1.5 × 108 CFU/mL at hub of needleless connector. Then, the solution obtained by washing each hub with sterile water was smeared onto the medium. After the culture, no colonies were observed after 2 hours. The results of this study confirmed that the antimicrobial activity of the disinfecting caps for needleless connectors is maintained for up to 7 days. These findings suggest that the clinical application of these caps can reduce the frequency of cap replacement, thereby shortening patient procedure times and contributing to a decrease in the incidence of bloodstream infections, providing fundamental data for their clinical implementation.
한국어
본 연구는 needleless connector에 부착하는 70% isopropyl alcohol 함유한 소독 캡의 항균 활성 및 지속성을 확인하고자 진행하였다. Needleless connector의 hub에 Staphylococcus aureus , Staphylococcus epidermidis, Escherichia coli , Pseudomonas aeruginosa, Acinetobacter baumannii, Candida albicans, Candida glabrata를 1.5 × 108 CFU/mL 접종 후 1분, 2시간, 4시간 및 7일 동안 실온에 방치하였다. 각 hub의 멸균수 세척액을 배지에 배양하였으며, 2시간 이후부터는 모든 실험 균주의 집락이 관찰되지 않았다. 본 연구 결과 needleless connector용 소독 캡의 항균 활성이 7일까지 유지됨을 확인하였으며, 이는 임상 환경에서 소독 캡의 교체 빈도를 감소시켜 환자 처치 시간을 단축시키고, 혈류감염 발생률 감소에 기여함으로써 임상 적용 가능성에 대한 기초자료를 제공한다.
목차
요약 Abstract 1. 서론 2. 연구 방법 2.1 연구대상 2.2 실험균주와 배지 2.3 현탁액 만들기 2.4 실험균주 접종 및 배양 3. 연구 결과 4. 고찰 5. 결론 REFERENCES