The Effect of Backscattered Radiation on Image Quality in Radiography Using Software-based Scatter Correction
Software-based Scatter Correction을 이용한 X선 검사 시 후방 산란선이 영상 화질에 미치는 영향
This study investigated the effect of BSR on image quality in radiography using SW-SC technique. A human phantom was used to acquire lateral lumbar images under 48 different geometric conditions by varying SID, X-ray exposure condition, collimation size, phantom thickness and rear shielding. Image quality was evaluated using the CIF, SSIM Index and qualitative assessment. The results demonstrated that image degradation due to BSR was prominent under conditions of long SID, low exposure, wide collimation, thick phantoms and the absence of rear shielding. Among these variables, phantom thickness demonstrated the strongest association with BSR-induced image degradation. In conclusion, when applying SW-SC techniques in clinical radiography, the use of rear shielding is particularly recommended for patients with larger body habitus to effectively minimize BSR-related degradation of image quality.
한국어
본 연구는 SW-SC 기법을 사용한 X선 검사 시 BSR이 영상 화질에 미치는 영향을 평가하였다. 인체 팬텀을 사용하여 SID, X선 노출조건, 조사야 크기, 피사체의 두께, 후방 차폐체 유무를 조합한 48가지 기하학적 조건에서 요추 측면 영상을 획득하였다. 영상 분석을 위해 CIF, SSIM Index, 정성적 평가를 진행하였다. 연구 결과, 긴 SID, 낮은 노출조건, 넓은 조사야, 두꺼운 피사체, 후방 차폐체가 없는 조건에서 BSR로 인한 영상 화질 저하가 발생하였으며, 특히 피사체의 두께는 BSR 발생과 가장 밀접한 관련이 있었다. 결론적으로, SW-SC 기법을 이용한 X선 검사 시 체격이 큰 환자에게는 후방 차폐체를 사용하는 것이 영상의 화질 저하를 최소화하는 효과적인 방법일 것이다.