In this study, we evaluated image quality applying median filter which is one of the noise reduction methods in the degraded nuclear medicine imaging. To acquire the PET/MR imaging, the mMR which is PET/MR scanner (Siemens, Germany) was used. The Jaszczak phantom was filled with 18F radioisotope during 10 minutes applying ultrashort echo time (UTE) magnetic resonance imaging (MR) pulse sequence as attenuation correction. To evaluate the image quality, we applied the noise reduction filter such as Wiener and median filter in the degraded images, respectively. In addition, we used contrast to noise ratio (CNR) and coefficient of variation (COV) by setting the region of interests. As the results, the image quality with median filter is superior than images with noise, wiener, and median filters through CNR and COV. In conclusion, we confirmed that image quality was improved when the median filter which is one of the noise reduction filters was used.
한국어
본 연구의 목적은, ultrashort echo time (UTE) magnetic resonance imaging (MRI) 펄스 시퀀스를 적용한 positron emission tomography (PET) /MRI 영상의 Wiener filter 와 median 필터를 적용하여 노이즈 감소 정도를 평가하였다. 영상 평가를 위해 18F 솔루션을 Jaszczak 팬텀에 주입하고 10분 동안 PET/MRI 영상을 획득하였다. 획득 영상에 Gaussian 노이즈를 부가하여 Wiener filter 와 median filter를 적용하여 contrast to noise ratio (CNR) 과 coefficient of variation (COV) 의 정량적 평가인자를 적용하여 영상을 평가하였다. CNR 과 COV의 결과 값을 기반으로 median 필터를 적용한 PET/MRI 영상이 우수함을 확인할 수 있었다.
목차
Abstract Ⅰ. 서론 Ⅱ. 재료 및 방법 1. 실험대상 및 설계 2. 정량적 평가 Ⅲ. 결과 Ⅳ. 고찰 및 결론 참고문헌 요약
키워드
Positron emission tomography (PET) / magnetic resonance imaging (MRI)Nuclear medicine imageImage qualityPositron emission tomography (PET) / magnetic resonance imaging (MRI)Nuclear medicine imageImage quality
저자
Chan Rok Park [ 박찬록 | Department of Radiological Science, Jeonju University ]
Corresponding Author