초록
영어
In this study, we evaluated image quality for the cadmium zinc telluride (CZT) semiconductor based gamma caemra compared with NaI(Tl) scintillator detector. We modeled gamma camera based on the CZT and NaI(Tl), and spatial resolution phantom which consists of various diameters. The phantom was filled with 99mTc in 140 keV and acquired during 500 s using GATE simulation tool. To analysis the image performance, the region of interest was set in the simulated image. The contrast to noise ratio (CNR) was used. According to the results, the CNR results for the NaI(Tl) and CZT show the 0.66 and 0.81, respectively. Compared with CZT and NaI(Tl) materials, image quality for CNR is improved approximately 18 % when the CZT based gamma camera was applied. We confirmed that CZT based gamma camera is effective to improve the image quality in nuclear medicine imaging system.
한국어
본 연구의 목적은, Cadmium Zinc Telluride (CZT) 반도체 기반 감마카메라 영상의 유용성을 평가해보고자 한다. geant4 application for tomographic emission (GATE) 시뮬레이션 툴을 이용해 NaI(Tl) 섬광체 와 CZT 반도체 물질 기반 감마카메라와 공간 분해능 평가 팬텀을 각각 모델링하였다. 140 keV 99mTc 의 방사성 동위원소를 주입하여 500 초 동안 영상을 획득 하고, 획득 영상의 관심영역을 설정하여 영상을 평가하였다. 영상평가를 위해 contrast to noise ratio (CNR) 정량평가 분석 방법을 사용하였고, 2.1 mm 지름을 가진 구와 background 에 관심영역을 각각 설정하였다. 결과적으로 반도체 기반 CZT 감마카메라 영상이 약 18% 향상됨을 확인할 수 있었다.
목차
Abstract Ⅰ. 서론 Ⅱ. 재료 및 방법 1. 감마카메라와 공간 분해능 팬텀 모델링 2. 정량적 분석 Ⅲ. 결과 Ⅳ. 고찰 및 결론 참고문헌 요약
키워드
Geant4 application for tomographic emission (GATE)Semiconductor detectorGamma cameraMonte Carlo simulationGeant4 application for tomographic emission (GATE)Semiconductor detectorGamma cameraMonte Carlo simulation
저자
Chan Rok Park [ 박찬록 | Department of Nuclear Medicine, Seoul National University Hospital ]
Corresponding Author