Monte Carlo simulation study using quadrant bar phantom in pixelated CdTe semiconductor gamma camera system
몬테 카를로 시뮬레이션 기반 Quadrant bar 팬텀과 γ-camera 모델링을 통한 영상 획득 및 평가
※ 기관로그인 시 무료 이용이 가능합니다.
※ 학술발표대회집, 워크숍 자료집 중 4페이지 이내 논문은 '요약'만 제공되는 경우가 있으니, 구매 전에 간행물명, 페이지 수 확인 부탁 드립니다.
4,000원
원문정보
초록
영어
In this study, we modeled gamma camera with cadmium telluride (CdTe) pixelated semiconductor detector using Geant4 application tomographic emission simulation tool using the quadrant bar phantom and evaluated image quality as visualization evaluation. We modeled CdTe semiconductor detector and pixelated parallel hole collimator with same pixel size as pitch size. The quadrant bar phantom was also simulated with four bars which consists of different length (0.5, 1, 1.5 and 2 mm). The radioistope and activity were used by 99mTc, 100 kBq during 500 s, respectively. Based on the visualization evaluation, it is confirmed that 1.5 and 2 mm bar conditions is distinguishable. In conclusion, we achieved excellent result by modeling the gamma camera system using the GATE simulation tool.
한국어
본 연구의 목적은, 핵의학 분야에서 주로 이용되는 Geant4 application tomographic emission (GATE) 시뮬레이 션을 툴을 사용하여 quadrant bar 팬텀 과 γ-camera를 모사하여 영상을 획득하고, 육안평가를 통해 영상의 질을 비교 하였다. cadmium telluride 반도체 디텍터와 평행다공형 콜리메이터를 사용해 영상을 획득하였고, 4 분할로 구성되어 있는 bar 팬텀 (0.5, 1, 1.5, 2 mm)을 GATE 시뮬레이션 툴을 이용해 방사성동위원소 99mTc, 100 kBq 의 용량으로 500 초 동안 영상을 획득 했다. 획득한 영상을 육안으로 평가했을 때, 1.5 와 2 mm bar 영상에서 각 bar 별 구별이 가능함을 확인 할 수 있었다. 본 연구를 통해 quadrant bar 팬텀 과 γ-camera를 완벽하게 모사하였고, 영상획득 후 평가하였다.
목차
Abstract Ⅰ. 서론 Ⅱ. 재료 및 방법 1. γ-camera 설계 2. γ-camera 설계 Ⅲ. 결과 Ⅳ. 고찰 및 결론 참고문헌 요약
키워드
Geant4 Application Tomographic Emission(GATE)Nuclear medicine imageImage qualityGeant4 application tomographic emission (GATE)Nuclear medicine imageImage quality
저자
Chan Rok Park [ 박찬록 | Department of Nuclear Medicine, Seoul National University Hospital ]
Corresponding Author