흉부 Phantom을 이용한 Low Dose CT의 관전압과 ASIR(Adaptive Statistical Iterative Reconstruction)적용에 따른 영상평가 및 피폭선량에 관한 연구
A study of image evaluation and exposure dose with the application of Tube Voltage and ASIR of Low dose CT Using Chest Phantom

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 16 Number 2 (2014.12)바로가기
  • 페이지
    pp.9-14
  • 저자
    황혜성, 김누리, 정윤지, 구은회, 김기정
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A483074

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
Purpose: The purpose of this study has attempted to evaluate and compare the image evaluation and exposure dose by respectively applying Filtered Back Projection(FBP), the existing test method, and Adaptive Statistical Iterative Reconstruction(ASIR) with different values of tube voltage during the Low Dose Computed Tomography(LDCT). Materials and Methods: With the image reconstruction method as basis, Chest Phantom was utilized with the FBP and ASIR set at 10%, 20% respectively, and the change of Tube Voltage (100kVp, 120kVp). For image evaluation, Back ground noise, Signal to Noise ratio(SNR) and Contrast to Noise ratio(CNR) were measured, and, for dose evaluation, CTDIvol and DLP were measured respectively. The statistical analysis was tested with SPSS(ver. 22.0), followed by ANOVA Test conducted after normality test and homogeneity test. (p<0.05). Results: In terms of image evaluation, there was no outstanding difference in Ascending Aorta(AA) SNR and Infraspinatus Muscle(IM) SNR with the different values of ASIR application(p<0.05), but a significant difference with the different amount of tube voltage(p>0.05). Also, there wasn't noticeable change in CNR with ASIR and different amount of Tube Voltage (p<0.05). However, in terms of dose evaluation, CTDIvol and DLP showed contrasting results(p<0.05). In terms of CTDIvol, the measured values with the same tube voltage of 120kVp were 2.6mGy with No-ASIR and 2.17mGy with 20%-ASIR respectively, decreased by 0.43mGy, and the values with 100kVp were 1.61mGy with No-ASIR and 1.34mGy with 20%-ASIR, decreased by 0.27mGy. In terms of DLP, the measured values with 120kVp were 103.21mGy·cm with No-ASIR and 85.94mGy·cm with 20%-ASIR, decreased by 17.27mGy·cm(about 16.7%), and the values with 100kVp were 63.84mGy·cm with No-ASIR and 53.25mGy·cm with 20%-ASIR, a decrease by 10.62mGy·cm(about 16.7%). Conclusion: At lower tube voltage, the rate of dose significantly decreased, but the negative effects on image evaluation was shown due to the increase of noise. For the future, through the result of the experiment, it is considered that the method above would be recommended for follow-up patients or those who get health checkup as long as there is no interference on the process of diagnosis due to the characteristics of Low Dose examination.

목차

Abstract
Ⅰ. 서론
Ⅱ. 대상 및 방법
2.1 대상 및 방법
2.2 영상획득 및 평가방법
2.3 영상분석(Image analysis)
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌

키워드

Chest Phantom Low dose CT ASIR

저자

  • 황혜성 [ Hyeseong Hwang | 청주대학교 방사선학과 ]
  • 김누리 [ Nuri Kim | 청주대학교 방사선학과 ]
  • 정윤지 [ Yoonji Jeong | 청주대학교 방사선학과 ]
  • 구은회 [ Eunhoe Goo | 청주대학교 방사선학과 ] Corresponding Author
  • 김기정 [ Kijeong Kim | 건국대학교병원 ]

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 16 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장