복부투시조영 검사 시 Added filter와 Grid 변화에 따른 선량 및 화질에 관한 연구
Study on dose and image quality by Added filter and Grid change when exam abdominal fluoroscopy

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 14 Number 2 (2012.10)바로가기
  • 페이지
    pp.47-56
  • 저자
    홍선숙, 강경미, 성민숙, 이종웅
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A483039

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
Amount of radiation exposure by seeing through fluoroscopy examination while is many patient exposure administration and unprepared misgovernment be. In this study, abdominal fluoroscopy during the scan, the dose and image quality change according to the use of grid and added filter optimized by measuring the test condition is proposed. Uses seeing through fluoroscopy examination equipment of Image Intensifier of Easy Diagnost Eleva (Philips), under tube type and uses Human phantom and measures average area dose according to grid insertion existence and nonexistence and added filter kind change. Measure sum of 29 organ dose and effective dose through PCXMC imagination simulation program and image J program through noise, SNR, image distortion was measured. Resolution, sharpness, and analyzed using the MTF curves. Fluorography the grid to insert the filter thickness and thickening and increased the average area dose and organ doses and effective dose. In the case of spot examination, when inserted grid, average area dose and organ dose and effective dose increased. Filter thickens the average area dose decreased, but the organ doses and effective dose were increased when use 0.2mmCu+1mmAl filter, decreased slightly. Noise and SNR measurements without inserting the gird, if you do not use the added filter was the lowest and when measure the distortion, 0.1mmCu+1mmAl filter was no difference of image quality in case insert grid was judged that when did not use occasion added filter that do not use grid, difference of image quality does not exist. Did not show a big difference, according to the grid and uses of the added filter sharpness, and resolution. Patient dose increases with factors that reduce the quality of the image so reckless grid and the use of the added filter when abdominal fluoroscopy examination should be cautious in using.

목차

Abstract
Ⅰ. 서론
Ⅱ. 대상 및 방법
1. Fluorography와 spot 검사의 면적선량 평가
2. 가상시뮬레이션 프로그램을 통한 장기선량및 유효선량 평가
3. Spot 영상의 noise 및 SNR 평가
4. Spot 영상의 SNR, PSNR 비교를 통한 화질 평가
5. Modulation Transfer Function (이하 MTF)곡선을 통한 spot 영상의 해상력 평가
Ⅲ. 결과
1. Fluorography와 spot 검사의 면적선량 평가분석
2. 가상시뮬레이션 프로그램을 통한 장기선량및 유효선량 평가 분석
3. Spot 영상의 noise 및 SNR 평가 분석
4. Spot 영상의 SNR, PSNR 비교를 통한 화질평가 분석
5. Modulation Transfer Function (이하 MTF)곡선을 통한 spot 영상의 해상력 화질 평가분석
Ⅳ. 고찰
Ⅴ. 결과
참고문헌

키워드

Abdominal fluoroscopy grid added filter dose image quality

저자

  • 홍선숙 [ Seon Sook Hong | 서울아산병원 영상의학팀 ] Corresponding Author
  • 강경미 [ Kyeong Mi Kang | 서울아산병원 영상의학팀 ]
  • 성민숙 [ Min Suk Seong | 서울아산병원 영상의학팀 ]
  • 이종웅 [ Jong Woong Lee | 강동경희대학교병원 영상의학과 ]

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 14 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장