C-arm을 이용한 EVAR(Endovascular Aneurysm Repair) 시술시 Lead Curtain 설치에 따른 Staff의 피폭선량 감소에 대한 연구
The investigation of the exposure dose reduction of the Staff according to the Lead Curtain installation in EVAR(Endovascular Aneurysm Repair) surgical operation using C-arm

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 14 Number 2 (2012.10)바로가기
  • 페이지
    pp.33-38
  • 저자
    유인웅, 정재연, 이관섭
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A483037

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to. The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result The height 3 the spot (50cm, 100cm, 150cm) dose was measured on 2 locations for 600sec in the X-ray radiation considering the surgical staff's movement. To install the curtains, we compared before and after the dose. As a result, it can confirm that dose of the installation former of 50cm height and after is reduced about 75% and 91% in 2 locations. In 100cm height, the reduction of the dose was a bit confirmed. There as to dose value, measured on 150cm height the installation former and after was nearly no change. This research examined the exposure dose about the radiation of the surgical staff during EVAR procedure in which the operation time is the long time. It was implemented in the object that it reduces the radiation exposure. It could confirm the certain effect of the experimental result exposure dose reduction.exposure dose reduction In EVAR procedure using long time C-arm, we studied exposure dose reduction and effeciency through measuring surgical staff's ESD by installing lead curtain operating table next to.

목차

Abstract
Ⅰ. 서론
Ⅱ. 연구대상 및 방법
1. 연구대상
2. 실험기기
3. 실험방법
Ⅲ. 결과
1. Lead curtain 설치 전과 후의 측정선량비교.
2. 측정높이에 따른 선량변화의 차이
Ⅳ. 고찰 및 결론
참고문헌

저자

  • 유인웅 [ In Woung Yoo | 서울아산병원 영상의학팀 ] Corresponding Author
  • 정재연 [ Jea Yeon Chung | 서울아산병원 영상의학팀 ]
  • 이관섭 [ Kwan Seob Lee | 서울아산병원 영상의학팀 ]

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 14 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장