디지털 환경에서 Humerus 검사 시 촬영인자 변화에 따른 선량 및 화질 평가
Dose and Image Assessment according to Radiologic Factors Variation at Digital Humerus X-ray Examination

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 14 Number 2 (2012.10)바로가기
  • 페이지
    pp.1-8
  • 저자
    김성민, 홍선숙, 이관섭, 하동윤
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A483033

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
Purpose : We aim at presenting the optimum radiologic factor through the evaluation of dose variation and of image quality through the use of a grid in Humerus examination and the change of dose because of the change of radiologic factor. Materials and Methods : We divided it in 3 cases: when using a grid or not and when using IP(Image Plate) in a digital system. Also, as fixing kVp to 70kVp it changed mAs, and fixing mAs to 10 it changed kVp, we put up resolution chart and Burger rose phantom on the acrylic phantom of 7cm (the same level of Humerus) to evaluate the dose and image. We used Image J program to evaluate the quantitative resolution of the obtained image, and made the qualitative evaluation and statistical analysis of the image saved in PACS for 20 radiologic technologist with more than 10 years of experience in order of evaluate its contrast. We used SPSS10(SPSS Inc. Chicago, Illinois) for statistical analysis. Results : We observed the analytic result of resolution by the change of kVp that it was 4.539dGycm2 in 60kVp and 757.472dGycm2 in 75kVp, which increased about 64.6% of dose, while for the resolution it had the pixel value 30.7% better with 851 in 60kVp than 651 in 75kVp. Also, we analyzed the result of resolution by the change of mAs that it was 3.106dGycm2 in 5mAs, and 12.470dGycm2 in 20mAs, which increased about 400% of dose, while for the resolution DR had 678 in 5mAs, and 724 in 20mAs that incr-eased about 6.8% of resolution. We made the qualitative evaluation of contrast by the change of kVp that DR showed the higher quality than CR, but the contrast by the change of kVp had no special different at the moment of visual evaluation, nor statistically significant difference(P>0.05). We observed the qualitative evaluation of contraste by the change of mAs that the contrast increased as DR increased mAs, and had statistically significant difference(P<0.05). On the other hand, CR had no significant difference for more than 10mAs nor statistically significant difference(P>0.05). Conclusion : In case of some patients with radiographic exposure by the repeated examination such as emergent patient or Follow up patient, they are considered to try to limit the use of a grid, to set kVp under 65kVp in fixed mode, to select less than 10mAs and to reduce the possibility of patient being bombed.

목차

Abstract
Ⅰ. 서론
Ⅱ. 실험기기 및 방법
1. 실험장비 및 측정기기
2. 실험방법
Ⅲ. 결과
1. 해상력 차트를 이용한 선량 및 Plot-profile분석 결과
2. Burger rose phantom을 이용한 대조도 정성적 평가
Ⅳ. 고찰
Ⅴ. 결론
참고문헌

키워드

Humerus examination Radiologic factors Burgerose phantom

저자

  • 김성민 [ Seong Min Kim | 서울아산병원 영상의학과 ] Corresponding Author
  • 홍선숙 [ Seon Sook Hong | 서울아산병원 영상의학과 ]
  • 이관섭 [ Kwan Sup Lee | 서울아산병원 영상의학과 ]
  • 하동윤 [ Dong Yun Ha | 서울아산병원 영상의학과 ]

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 14 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장