The purpose of this study was to survey and test quality control of mammography system. The conclusion of this study is as follows ; First, The rate of pass for phantom image test shows that Film-Screen mammography system(F/S) and computed mammography system(CR) is 80%, Indirect digital mammography system(DR) is 100%. Second, The test of exposure dose shows that F/S is 921 mR. CR is 1,140 mR, DR is 474 mR. The grade of this testament is CR > F/S > DR. Third, The test of average glandular dose shows that F/S is 1,336 mGy, CR is 1,635 mGy, DR is 1,26 mGy. The grade of this testament is CR > F/S > DR. Fourth, The testament of resolution shows as follows F/S is 11~13 Lp/mm, CR is 4~5 Lp/mm, and DR is 5~7 Lp/mm(F/S > DR > CR) Fifth, The survey of projection, cassette, development and reading shows that user are indifference.
목차
Abstract Ⅰ. 서론 Ⅱ. 연구 방법 1. F/S와 CR, DR system의 사용 실태 파악 2. F/S와 CR, DR system의 ACR 공인 phantom과 지방 phantom 영상에서 화질 측정 3. F/S와 CR, DR system의 피폭선량 측정 4. F/S와 CR, DR system의 해상력 측정 Ⅲ. 연구 결과 1. F/S와 CR, DR system의 사용 실태 파악 2. F/S와 CR, DR system의 ACR과 지방 phantom영상에서 화질 측정 3. F/S와 CR, DR system의 피폭선량 측정 4. F/S와 CR, DR system의 해상력 측정 5. F/S와 CR, DR system에서 촬영과 카세트, 현상 그리고 판독 설문조사 Ⅳ. 고찰 Ⅴ. 결론 참고문헌