재구성 알고리즘 변화에 따른 CT 영상의 화질 평가
The Evaluation of Image Quality According to the Change of Reconstruction Algorithm of CT Images

첫 페이지 보기
  • 발행기관
    대한디지털의료영상학회 바로가기
  • 간행물
    대한디지털의료영상학회논문지 바로가기
  • 통권
    Volume 12 Number 2 (2010.11)바로가기
  • 페이지
    pp.127-132
  • 저자
    한동균, 박건진, 고신관
  • 언어
    한국어(KOR)
  • URL
    https://www.earticle.net/Article/A482994

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
In this study, the correlation among the changes of Modulation Transfer Function(MTF) in the noise and high-contrast resolution and the change of Contrast to noise ratio(CNR) in the low-contrast resolution will be examined to investigate the estimation of image quality according to the type of algorithms. The image data obtained by scanning American Association of Physicists in Medicine(AAPM) phantom was applied to each algorithm and the exposure condition of 120 kVp, 250 mAs, and then the CT number and noise were measured. The MTF curved line of the high-contrast resolution was calculated with Point Spread Function(PSF) by using the analysis program by Philips, resulting in 0.5 MTF, 0.1 MTF and 0.02 MTF respectively. The low-contrast resolution was calculated with CNR and the uniformity was measured to each algorithm. Since the measurement value for the uniformity of the equipment was below ± 5 HU, which is the criterion figure, it was found to belong to the normal range. As the algorithm got closer from soft to edge, the standard deviation of CT number increased, which indicates that the noise increased as well. As for MTF, 0.5 MTF, 0.1 MTF and 0.02 MTF were all sharp algorithms, and as the algorithm got closer from soft to edge, it was possible to distinguish more clearly with the naked eye. On the other hand, CNR gradually decreased, because the difference between the contrast hole CT number and the acrylic CT number was the same while the noise of hole increased.

목차

Abstract
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌

키워드

CT image quality MTF Noise CNR

저자

  • 한동균 [ Dong Kyoon Han | 을지대학교 보건과학대학 방사선학과 ]
  • 박건진 [ Kun Jin Park | 삼성서울병원 영상의학과 ]
  • 고신관 [ Shin Kwan Ko | 을지대학교 보건과학대학 방사선학과 ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한디지털의료영상학회 [대한디지털의료영상학]
  • 설립연도
    1995
  • 분야
    의약학>방사선과학
  • 소개
    대한디지털의료영상학회는 디지털 영상 기술에 관한 연구와 실험 결과를 임상에 적용하여 영상의학 및 진단 방사선 기술 분야의 발전에 기여함을 목적으로 한다.

간행물

  • 간행물명
    대한디지털의료영상학회논문지 [Korean Journal of Digital Imaging in Medicine]
  • 간기
    연3회
  • pISSN
    1976-7641
  • 수록기간
    1995~2025
  • 십진분류
    KDC 510 DDC 610

이 권호 내 다른 논문 / 대한디지털의료영상학회논문지 Volume 12 Number 2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장