초록
영어
This experimental study is carried out one of the General Hospital in Kyungbok providence. Abdomen Phantom being located Anterior-posterior(AP) position on portable bed, and the portable X-ray generating device was placed the phantom at -90° direction. The experiment were set 65 kVp, 10 mAs, 10 × 10 cm2, 100 cm(FOD) for the measurement. Digital proportional counting tube survey meter was used for measuring the space scatter dose. Measurement points of horizontal distribution was set up at 30° interval by increasing 50 cm radius of upside, downside, left and right. Vertical distribution of measurement points were set up for the vertical plane with a radius of at 30° intervals with 50cm increments. It is concluded that longer distance from the soure of X-ray significanly decrease radiation dose to the patient and use of the radiation protection device should be applied in clinical practice to reduce dose to the patient.
목차