영어
Background: Accurate quantification of surface dose is critical in radiation therapy, particularly for patients with breast cancer, for whom the risk of cutaneous toxicity is substantial. In this study, we evaluated the clinical feasibility of a novel flexible radiochromic film specifically designed to improve adhesion and usability on curved anatomic surfaces. Materials and Methods: The study prospectively enrolled 20 patients with breast cancer undergoing volumetric modulated arc therapy. Surface dose was measured using flexible radiochromic film placed near the surgical scar and close to the isocenter. Calibration curves were established using a 6 MV photon beam with flattening filter free and without, and the film response was analyzed in both transmission and reflection scanning modes. The measured surface doses were then compared with those calculated by the treatment planning system (TPS). Results and Discussion: The flexible radiochromic film demonstrated excellent sensitivity in the 1.5–2.5 Gy dose range, with superior performance observed in the transmission scanning mode. The film adhered securely to skin without requiring additional fixation and was removed without causing discomfort to the patient. Notably, substantial discrepancies were found between the measured and TPS-calculated surface doses, underscoring the limitations of a TPS in accurately modeling skin dose. Conclusion: Flexible radiochromic film demonstrated dosimetric performance similar to that of conventional films while offering enhanced flexibility and ease of clinical use. Given the significant uncertainty in TPS-based surface-dose estimation, the film provides accurate and practical in vivo surface-dose assessment in breast cancer radiotherapy.

목차

ABSTRACT
Introduction
Materials and Methods
1. Patient Selection
2. Film Processing (Calibration, Scanning, and Analysis)
3. Measuring In Vivo Surface Dose
4. Comparing Film Results with TPS Estimates
5. Statistical Analysis
Results and Discussion
Conclusion
Article Information
References

키워드

Flexible Film Radiochromic Film Breast Cancer Skin Dose Surface Measurement

저자

  • Jin Jegal [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea ]
  • Jin Dong Cho [ Department of Radiation Oncology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea ]
  • Jung-In Kim [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea ]
  • Chang Heon Choi [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea ]
  • Bum-Sup Jang [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea ]
  • Jin Ho Kim [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea ]
  • Kyung Hwan Shin [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea; Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea; Department of Radiation Oncology, Seoul National University College of Medicine, Seoul, Republic of Korea ]
  • So-Yeon Park [ Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea; Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Republic of Korea ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한방사선방어학회 [Korean Association For Radiation Protection]
  • 설립연도
    1975
  • 분야
    자연과학>기타자연과학
  • 소개
    회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다. 1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최 2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포 3. 방사선방어에 관한 학술의 국제교류 및 협력 4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역 5. 방사선방어에 관한 조사 및 연구용역 6. 회원의 연구활동을 위한 제반협조 7. 기타 본 학회의 목적 달성에 필요한 사항

간행물

  • 간행물명
    방사선방어학회지 [Journal of Radiation Protection and Research]
  • 간기
    계간
  • pISSN
    2508-1888
  • 수록기간
    1976~2026
  • 등재여부
    KCI 등재,SCOPUS
  • 십진분류
    KDC 559 DDC 629

이 권호 내 다른 논문 / 방사선방어학회지 VOLUME 50 NUMBER 4

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장