Background: Vaginal cuff brachytherapy after hysterectomy targets radiation to potential residual cancer cells. Our previous study proposed a three-ovoid applicator to improve dose coverage and stability over conventional applicators but was limited to two-dimensional plan comparisons. This study introduces a polymethyl methacrylate (PMMA)-Agar phantom for threedimensional (3D) dosimetric comparison of these applicators and verifies its applicability. Materials and Methods: Cylinder, two-ovoid, and three-ovoid applicators were compared. A PMMA frame (30 cm×20 cm×10 cm) with a central 10 cm diameter cavity for the Agar phantom was constructed. Using treatment plans delivering 3 Gy to a 5 mm thick bolus target, dose distributions were verified using Gafchromic EBT3 film (Ashland Inc.) and an Exradin A1SL thimble ionization chamber (Standard Imaging Inc.). Results and Discussion: Plan evaluations showed the three-ovoid applicator demonstrated improved dose uniformity and target coverage over conventional applicators. It yielded higher mean target doses (1.97 Gy vs. cylinder; 0.78 Gy vs. two-ovoid normal) and notably enhanced dose homogeneity, reducing a key uniformity index (D5 cm³/D99 Vol%) from 2.08 (cylinder) and 1.68 (two-ovoid normal) down to 1.61. These planned advantages were qualitatively supported by A1SL chamber and EBT3 film measurements showing dose distributions consistent with plans, experimentally verifying the three-ovoid’s dosimetric advantage using the PMMA-Agar phantom. Conclusion: In this study, we experimentally verified the three-ovoid applicator’s dosimetric advantage compared to the conventional applicators by using an in-house designed PMMAAgar phantom for vaginal cuff brachytherapy. We plan to improve the experimental 3D dose verification accuracy in the future, considering the more realistic patient’s internal organ structure using a 3D printing technique.

목차

ABSTRACT
Introduction
Materials and Methods
1. Applicators for Vaginal Cuff Brachytherapy
2. In-house Designed PMMA-Agar Phantom
3. Brachytherapy Planning
4. Dose Measurement
Results
1. Phantom Quality Evaluation
2. Plan Quality Evaluation for Various Applicators
3. Measured Dose Evaluation
Discussion
Conclusion
Article Information
References

키워드

Vaginal Cuff Brachytherapy High Dose Rate PMMA-Agar Phantom 3D Dose Verification Three-Ovoid Dose Measurement

저자

  • Saerom Sung [ Department of Radiation Convergence Engineering, Yonsei University, Wonju, Republic of Korea ]
  • Hyemi Kim [ Department of Radiological Science, Jeonju University, Jeonju, Republic of Korea ]
  • Sunghyun Kim [ Department of Radiation Oncology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea ]
  • Misun Kim [ Department of Radiation Oncology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea ]
  • Hyejung Cha [ Department of Radiation Oncology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea ]
  • Yeon Soo Yeom [ Department of Radiation Convergence Engineering, Yonsei University, Wonju, Republic of Korea ]
  • Sei Hwan You [ Department of Radiation Oncology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea ]
  • Chul Hee Min [ Department of Radiation Convergence Engineering, Yonsei University, Wonju, Republic of Korea ] Corresponding Author
  • Hyun Joon Choi [ Department of Radiation Oncology, Wonju Severance Christian Hospital, Yonsei University Wonju College of Medicine, Wonju, Republic of Korea ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한방사선방어학회 [Korean Association For Radiation Protection]
  • 설립연도
    1975
  • 분야
    자연과학>기타자연과학
  • 소개
    회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다. 1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최 2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포 3. 방사선방어에 관한 학술의 국제교류 및 협력 4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역 5. 방사선방어에 관한 조사 및 연구용역 6. 회원의 연구활동을 위한 제반협조 7. 기타 본 학회의 목적 달성에 필요한 사항

간행물

  • 간행물명
    방사선방어학회지 [Journal of Radiation Protection and Research]
  • 간기
    계간
  • pISSN
    2508-1888
  • 수록기간
    1976~2026
  • 등재여부
    KCI 등재,SCOPUS
  • 십진분류
    KDC 559 DDC 629

이 권호 내 다른 논문 / 방사선방어학회지 VOLUME 50 NUMBER 3

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장