Background: A portable gamma irradiation system was developed for on-site calibration of environmental radiation monitoring devices. This system uses a portable collimator and gammaray sources such as 57Co, 133Ba, 137Cs, and 60Co, each with a nominal radioactivity of 10 MBq or less. Materials and Methods: The air kerma rate produced by the system was measured using a spherical ionization chamber with a 10 L sensitive volume. This chamber was calibrated against the air kerma rate measured with primary standard ionization chambers. The system’s performance was evaluated by comparing the calibration results of the monitoring device with those obtained using a conventional method involving a standard ionization chamber for outdoor use. Results and Discussion: The calibration coefficient of the spherical ionization chamber was within ±2% of the value of 3.029×103 Gy/C for photon energies ranging from 50 keV to 1,250 keV. The air kerma rate of the irradiation field produced by the portable system ranged from 0.17 μGy/hr to 4.7 μGy/hr, with an uncertainty between 2.4% and 10%. The calibration coefficients obtained using the portable system were consistent with those from the conventional method. Conclusion: The portable gamma irradiation system offers significant advantages, including preventing damage to standard ionization chambers used outdoors and reducing measurement time compared to conventional methods. Due to its ease of construction and implementation, the system is expected to be useful not only for on-site calibration of monitoring devices but also in various radiation safety management scenarios.
목차
ABSTRACT Introduction Materials and Methods 1. Portable Gamma Irradiation System 2. Calibration of the Spherical Ionization Chamber 3. Calibration of the Irradiation Fields Produced with the Portable Gamma Irradiation System 4. On-site Calibration of the Monitoring Post Results and Discussion 1. Calibration Coefficient of the Spherical Ionization Chamber 2. Air Kerma Rate of the Gamma-Ray Field from the Irradiation System 3. Calibration Coefficient of the Monitoring Post Conclusion Conflict of Interest Acknowledgements Ethical Statement Author Contribution References
대한방사선방어학회 [Korean Association For Radiation Protection]
설립연도
1975
분야
자연과학>기타자연과학
소개
회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다.
1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최
2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포
3. 방사선방어에 관한 학술의 국제교류 및 협력
4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역
5. 방사선방어에 관한 조사 및 연구용역
6. 회원의 연구활동을 위한 제반협조
7. 기타 본 학회의 목적 달성에 필요한 사항
간행물
간행물명
방사선방어학회지 [Journal of Radiation Protection and Research]