Earticle

현재 위치 Home

Original Research

Impact of 0.35 T Magnetic Field on Dose Calculation for Non-small Cell Lung Cancer Stereotactic Radiotherapy Plans

첫 페이지 보기
  • 발행기관
    대한방사선방어학회 바로가기
  • 간행물
    방사선방어학회지 KCI 등재 SCOPUS 바로가기
  • 통권
    VOLUME 48 NUMBER 3 (2023.09)바로가기
  • 페이지
    pp.117-123
  • 저자
    Jaeman Son, Sung Young Lee, Chang Heon Choi, Jong Min Park, Jung-in Kim
  • 언어
    영어(ENG)
  • URL
    https://www.earticle.net/Article/A436208

원문정보

초록

영어
Background: We investigated the impact of 0.35 T magnetic field on dose calculation for nonsmall cell lung cancer (NSCLC) stereotactic ablative radiotherapy (SABR) in the ViewRay system (ViewRay Inc.), which features a simultaneous use of magnetic resonance imaging (MRI) to guide radiotherapy for an improved targeting of tumors. Materials and Methods: Here, we present a comprehensive analysis of the effects induced by the 0.35 T magnetic field on various characteristics of SABR plans including the plan qualities and dose calculation for the planning target volume, organs at risk, and outer/inner shells. Therefore, two SABR plans were set up, one with a 0.35 T magnetic field applied during radiotherapy and another in the absence of the field. The dosimetric parameters were calculated in both cases, and the plan quality indices were evaluated using a Monte Carlo algorithm based on a treatment planning system. Results and Discussion: Our findings showed no significant impact on dose calculation under the 0.35 T magnetic field for all analyzed parameters. Nonetheless, a significant enhancement in the dose was calculated on the skin surrounding the tumor when the 0.35 T magnetic field was applied during the radiotherapy. This was attributed to the electron return effect, which results from the deviation of the electrons ejected from tissues upon radiation due to Lorentz forces. These returned electrons re-enter the tissues, causing a local dose increase in the calculated dose. Conclusion: The present study highlights the impact of the 0.35 T magnetic field used for MRI in the ViewRay system for NSCLC SABR treatment, especially on the skin surrounding the tumors.

목차

ABSTRACT
Introduction
Materials and Methods
1. Patient Selection
2. Treatment System (ViewRay)
3. SABR Plans for Lung Cancer
4. Statistics Analysis
5. Plan Evaluation
Results and Discussion
1. PTV Dose Distribution Difference
2. Plan Qualities Difference
3. Organs at Risk Dose Distribution Difference
4. Dose Distribution Difference around the Skin
Conclusion
Conflict of Interest
Acknowledgements
Ethical Statement
Author Contribution
References

저자

  • Jaeman Son [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea ]
  • Sung Young Lee [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea ]
  • Chang Heon Choi [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea ]
  • Jong Min Park [ Department of Radiation Oncology, Seoul National University College of Medicine, Seoul, Korea ]
  • Jung-in Kim [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한방사선방어학회 [Korean Association For Radiation Protection]
  • 설립연도
    1975
  • 분야
    자연과학>기타자연과학
  • 소개
    회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다. 1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최 2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포 3. 방사선방어에 관한 학술의 국제교류 및 협력 4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역 5. 방사선방어에 관한 조사 및 연구용역 6. 회원의 연구활동을 위한 제반협조 7. 기타 본 학회의 목적 달성에 필요한 사항

간행물

  • 간행물명
    방사선방어학회지 [Journal of Radiation Protection and Research]
  • 간기
    계간
  • pISSN
    2508-1888
  • 수록기간
    1976~2026
  • 등재여부
    KCI 등재,SCOPUS
  • 십진분류
    KDC 559 DDC 629

이 권호 내 다른 논문 / 방사선방어학회지 VOLUME 48 NUMBER 3

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장