Background: The hemi-body electron beam irradiation (HBIe–) technique has been proposed for the treatment of mycosis fungoides. It spares healthy skin using an electron shield. However, shielding electrons is complicated owing to electron scattering effects. In this study, we developed a thimble-like head bolus shield that surrounds the patient’s entire head to prevent irradiation of the head during HBIe–. Materials and Methods: The feasibility of a thimble-like head bolus shield was evaluated using a simplified Geant4 Monte Carlo (MC) simulation. Subsequently, the head bolus was manufactured using a three-dimensional (3D) printed mold and Ecoflex 00-30 silicone. The fabricated head bolus was experimentally validated by measuring the dose to the Rando phantom using a metal-oxide-semiconductor field-effect transistor (MOSFET) detector with clinical configuration of HBIe–. Results and Discussion: The thimble-like head bolus reduced the electron fluence by 2% compared with that without a shield in the MC simulations. In addition, an improvement in fluence degradation outside the head shield was observed. In the experimental validation using the inhouse- developed bolus shield, this head bolus reduced the electron dose to approximately 2.5% of the prescribed dose. Conclusion: A thimble-like head bolus shield for the HBIe– technique was developed and validated in this study. This bolus effectively spares healthy skin without underdosage in the region of the target skin in HBIe–.
목차
ABSTRACT Introduction Materials and Methods 1. Monte Carlo Simulation 2. Experimental Configuration Results and Discussion Conclusion Conflicts of Interest Acknowledgements Author Contribution References
키워드
Hemi-Body Electron Beam IrradiationBolusElectron ShieldingMOSFET DetectorMonte Carlo Simulation
저자
Wook-Geun Shin [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; Biomedical Research Institute, Seoul National University Hospital, ]
Sung Young Lee [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea ]
Hyeongmin Jin [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Korea; Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, Korea ]
Jeongho Kim [ Department of Radiation Oncology Samsung Changwon Hospital, Sungkyunkwan University School of Medicine, Changwon, Korea ]
Seonghee Kang [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Korea; Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, Korea ]
Jung-in Kim [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Korea; Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, Korea ]
Seongmoon Jung [ Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; Biomedical Research Institute, Seoul National University Hospital, Seoul, Korea; Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, Korea ]
Corresponding Author
대한방사선방어학회 [Korean Association For Radiation Protection]
설립연도
1975
분야
자연과학>기타자연과학
소개
회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다.
1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최
2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포
3. 방사선방어에 관한 학술의 국제교류 및 협력
4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역
5. 방사선방어에 관한 조사 및 연구용역
6. 회원의 연구활동을 위한 제반협조
7. 기타 본 학회의 목적 달성에 필요한 사항
간행물
간행물명
방사선방어학회지 [Journal of Radiation Protection and Research]