Purpose: The purpose of this study was to evaluate the stability of abutment screws used with the zirconia fixture-based implant system and compare them with those used with the existing titanium fixture system via the finite element method. Methods: A single implant-supported restoration was designed for the finite element analysis. A universal analysis program was used to set 8 occlusal points along the direction to the long axis of the implant, and an occlusal load of 700 N was applied. Results: In all models (Zir and Ti-fixture model), the screw threads presented with the highest von Mises stress (VMS) values, whereas the head and end presented with the lowest VMS values. The VMS of the screw used in the zirconia-fixture model was 5.97% lower than that used in the titanium-fixture model (261.258 vs. 276.911 MPa, respectively) despite statistical significance. Furthermore, the zirconia fixture (352.912 MPa) had a higher stress value (8.42%) than the titanium fixture (332.331 MPa). In a completely tightened titanium fixture implant system, the stress was concentrated in the implant-abutment connection interface, the zirconia fixture presented with a stable stress distribution. Conclusion: Although the zirconia fixture demonstrated a high VMS value, owing to the stiffness and elasticity coefficients of the material, the stress generated in the abutment screws was similar in all models. In conclusion, the zirconia fixture-based implant system presented with a more stable stress distribution in the abutment screws than the titanium fixture-based implant system.

목차

Abstract
INTRODUCTION
MATERIALS AND METHODS
1. 유한 요소 모델
2. 경계 및 하중 조건
RESULTS
DISCUSSION
CONCLUSIONS
CONFLICT OF INTERESTS
ORCID
REFERENCES

키워드

Dental implants Finite element analysis Implant-supported dental prosthesis Zirconium oxide

저자

  • 김은영 [ Eun Young Kim | 부산가톨릭대학교 보건과학대학 치기공학과 ]
  • 홍민호 [ Min-Ho Hong | 부산가톨릭대학교 보건과학대학 치기공학과 ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한치과기공학회 [The Korean Academy Of Dental Technology]
  • 설립연도
    1979
  • 분야
    의약학>치의학
  • 소개
    본 학회는 치과기공에 관한 모든 분야의 연구와 회무를 총괄하기 위하여 세분화된 분과학회의 회원 상호간 학술교류와 정보를 수집하고, 보수교육 및 학술대회 준비, 진행과 학회지 발간등의 업무를 효율적으로 수행함을 목적으로 한다.

간행물

  • 간행물명
    대한치과기공학회지 [THE JOURNAL OF KOREAN ACADEMY OF DENTAL TECHNOLOGY]
  • 간기
    계간
  • pISSN
    1229-3954
  • eISSN
    2288-5218
  • 수록기간
    1979~2026
  • 등재여부
    KCI 등재
  • 십진분류
    KDC 515 DDC 617

이 권호 내 다른 논문 / 대한치과기공학회지 Vol.43 No.2

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장