Seung Kyu Lee, Insu Chang, Sang In Kim, Jungil Lee, Hyoungtaek Kim, Jang-Lyul Kim, Min Chae Kim
언어
영어(ENG)
URL
https://www.earticle.net/Article/A356538
원문정보
초록
영어
Background: In the calibration and testing laboratory of Korea Atomic Energy Research Institute, the old X-ray generator used for the production of reference X-ray fields was replaced with a new one. For this newly installed X-ray irradiation system, beam alignment as well as the verification of beam qualities was conducted. Materials and Methods: The existing X-ray generator, Phillips MG325, was replaced with YXLON Y.TU 320-D03 in order to generate reference X-ray fields. Theoretical calculations and Monte Carlo simulations were used to determine initial filter thickness. Beam alignment was performed in three steps to deliver a homogeneous radiation dosage to the target at different distances. Finally, the half-value layers were measured for different X-ray fields to verify beam qualities by using an ion chamber. Results and Discussion: Beam alignment was performed in three steps, and collimators and other components were arranged to maintain the uniformity of the mean air kerma rate within ±2.5% at the effective beam diameter of 28 cm. The beam quality was verified by using halfvalue layer measurement methods specified by American National Standard Institute (ANSI) N13.11-2009 and International Organization for Standardization (ISO)-4037. For each of the nine beams than can be generated by the new X-ray irradiation system, air kerma rates for Xray fields of different beam qualifies were measured. The results showed that each air kerma rate and homogeneity coefficient of the first and second half-value layers were within ±5% of the recommended values in the standard documents. Conclusion: The results showed that the new X-ray irradiation system provides beam qualities that are as high as moderate beam qualities offered by National Institute of Standards and Technology in ANSI N13.11-2009 and those for narrow-spectrum series of ISO-4037.
목차
ABSTRACT Introduction Materials and Methods 1. Establishment of reference X-ray field 2. Beam alignment of X-ray generator 3. Half-value layer measurement Results and Discussion Conclusion Acknowledgements References
대한방사선방어학회 [Korean Association For Radiation Protection]
설립연도
1975
분야
자연과학>기타자연과학
소개
회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다.
1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최
2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포
3. 방사선방어에 관한 학술의 국제교류 및 협력
4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역
5. 방사선방어에 관한 조사 및 연구용역
6. 회원의 연구활동을 위한 제반협조
7. 기타 본 학회의 목적 달성에 필요한 사항
간행물
간행물명
방사선방어학회지 [Journal of Radiation Protection and Research]