영어
Background: The mid-term performance of clinical linear accelerator (LINAC) during volumetric modulated arc therapy (VMAT) treatment period is not performed in clinical practice and usually replaced with one-time plan quality assurance (QA). In this research we aim to monitor daily reproducibility of VMAT delivery from tracking individual leaf movement error and dosimetric error to evaluate the mid-term quality of the machine used. Materials and Methods: First, multileaf collimator (MLC) information was imported into MATLAB program to determine which of the MLC leaves in the leaf bank had the maximum RMS position error (maxRMS). We estimated where the maximum positional errors (maxPE) of the chosen leaf occur along its path length and tracked its daily variations over the entire treatment period. Secondly, picture information of dosimetric error from portal dosimetry was imported into MATLAB where representative high gamma index region (HGR) was determined as HGR with length of > 1 cm and their centers were daily tracked. Results and Discussion: The maxPEs in the brain and tongue cases were distributed broader than in other cases, but all data were found located within ± 0.5 mm. From first day to last day all of five cases show the similar visual pattern of HGRs and Centers of the longest HGRs remained within ± 1 mm of that in first day. These findings prove excellent mid-term performance of the LINAC used in VMAT treatments over a full course of treatment. Conclusion: Tracking the daily location changes of leaf movement and dosimetric error can be a good indicator of predicting the daily quality like stability and reproducibility of beam delivering in VMAT treatment.

목차

ABSTRACT
Introduction
Materials and Methods
1. Equipment and tools
2. Selection of VMAT plans
3. Positional tracking of MLC movement error
4. Positional tracking of dosimetric error
Results and Discussion
1. Positional tracking of MLC movement error
2. Positional tracking of dosimetric error
Conclusion
ACKNOWLEDGEMENT
REFERENCES

키워드

VMAT Quality assurance MLC High gamma index Portal dosimetry

저자

  • Mohammad Mahfujur Rahman [ Department of Nuclear Engineering, Hanyang University, Seoul, Korea ]
  • Chan Hyeong Kim [ Department of Nuclear Engineering, Hanyang University, Seoul, Korea ]
  • Seonghoon Kim [ Department of Radiation Oncology, Hanyang University Medical Center, Hanyang University, Seoul, Korea ] Corresponding Author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    대한방사선방어학회 [Korean Association For Radiation Protection]
  • 설립연도
    1975
  • 분야
    자연과학>기타자연과학
  • 소개
    회원 상호간의 협조와 친목을 도모함으로써 방사선방어에 관한 제반연구 및 발전에 이바지함을 물론 학술의 국제교류 및 국제학술단체와의 상호협력 증진에 기여함을 목적으로 하며, 이 목적을 달성하기 위하여 다음 각 호의 사업을 한다. 1. 방사선방어에 관한 학술연구발표회 및 강연회 등의 개최 2. 학회지 및 방사선방어에 관한 학술간행물의 발행 및 배포 3. 방사선방어에 관한 학술의 국제교류 및 협력 4. 방사선방어에 관한 국제학술자료의 조사, 수집 및 번역 5. 방사선방어에 관한 조사 및 연구용역 6. 회원의 연구활동을 위한 제반협조 7. 기타 본 학회의 목적 달성에 필요한 사항

간행물

  • 간행물명
    방사선방어학회지 [Journal of Radiation Protection and Research]
  • 간기
    계간
  • pISSN
    2508-1888
  • 수록기간
    1976~2026
  • 등재여부
    KCI 등재,SCOPUS
  • 십진분류
    KDC 559 DDC 629

이 권호 내 다른 논문 / 방사선방어학회지 VOLUME 44 NUMBER 1

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장