The reproducibility and validity verification for body composition measuring devices using bioelectrical impedance analysis in Korean adults

첫 페이지 보기
  • 발행기관
    한국운동재활학회 바로가기
  • 간행물
    JER SCOPUS KCI 등재 바로가기
  • 통권
    Vol.14 No.4 (2018.08)바로가기
  • 페이지
    pp.621-627
  • 저자
    Seung-Won Yang, Tai-Hyung Kim, Hyun-Min Choi
  • 언어
    영어(ENG)
  • URL
    https://www.earticle.net/Article/A336639

※ 기관로그인 시 무료 이용이 가능합니다.

4,000원

원문정보

초록

영어
The purpose of this study is to examine the validity and reproducibility of impedance body fat measurement devices measuring the body com-position of Korean male and female adults using three bioelectrical im-pedance analyzers. We compared two methods for evaluating body composition: dual energy X-ray absorptiometry (DEXA), and bioelectri-cal impedance analysis (BIA). Subjects were 200 healthy adult Korean males and females whose mean±standard deviation (range) age, standing height, body weight, and body mass index (BMI) were 44.1±14.5 years, 172.8±7.4 cm, 76.0±12.8 kg, and 25.4±3.3 kg/m2, and 44.5±14.7 years, 158.7±5.8 cm, 58.3±8.3 kg, and 23.2±3.0 kg/m2, re-spectively. As a result, first of all, the reproducibility of the bioelectrical impedance analyzer had very high coefficients at r=0.998, r=0.997 be-tween men and women, respectively. The correlation coefficients among three comparisons for lean body mass (LBM) were provided the following coefficients: r=0.951 for DEXA vs. ACCUNIQ BC720, r=0.950 for DEXA vs. ACCUNIQ BC360, and r=0.946 for DEXA vs. ACCUNIQ BC380 in men. In the results for women, they also had the very high fol-lowing coefficients: r=0.956 for DEXA vs. ACCUNIQ BC720, r=0.946 for DEXA vs. ACCUNIQ BC360, and r=0.957 for DEXA vs. ACCUNIQ BC380 in LBM. In conclusion, this research showed a higher correlation in terms of accuracy compared to existing BIA-based body composition measurement techniques, and the accuracy of LBM was improved with high correlation coefficients through the algorithm that was improved using the multifrequency BIA method in the ACCUNIQ BC products.

목차

Abstract
 INTRODUCTION
 MATERIALS AND METHODS
  Survey subjects
  Experimental procedure
  Survey tool
  Data analyses
 RESULTS
  The characteristics of the research subjects
  The reproducibility verification for PBF and LBM using BIA
  The validity verification for LBM using DEXA and BIA
  The validity verification for PBF using DEXA and BIA
  The LBM and PBF validity verification using DEXA and BIA for men
  The LBM and PBF validity verification using DEXA and BIA for women
 DISCUSSION
 CONFLICT OF INTEREST
 ACKNOWLEDGMENTS
 REFERENCES

키워드

Validity Reproducibility Body composition Adults Bioelec-trical impedance analysis

저자

  • Seung-Won Yang [ Department of Physical Education, Baekseok University, Cheonan, Korea ]
  • Tai-Hyung Kim [ Graduate School of Technology Management, Kyung Hee University, Yongin, Korea ]
  • Hyun-Min Choi [ Department of Sports Leisure, Gwangju University, Gwangju, Korea ] Corresponding author

참고문헌

자료제공 : 네이버학술정보

간행물 정보

발행기관

  • 발행기관명
    한국운동재활학회 [Korean Society of Exercise Rehabilitation]
  • 설립연도
    2004
  • 분야
    의약학>재활의학
  • 소개
    한국운동재활학회는 사회적, 정신적, 신체적 통합건강복지 이론의 학술연구와 회원 상호간 학술교류 증진을 장려함으로써 학문적 발전을 도모하고 나아가 건강복지선진국 발전에 이바지함을 목적으로 한다.

간행물

  • 간행물명
    JER [Journal of Exercise Rehabilitation]
  • 간기
    격월간
  • pISSN
    2288-176X
  • eISSN
    2288-1778
  • 수록기간
    2013~2026
  • 등재여부
    SCOPUS,KCI 등재
  • 십진분류
    KDC 517 DDC 613

이 권호 내 다른 논문 / JER Vol.14 No.4

    피인용수 : 0(자료제공 : 네이버학술정보)

    함께 이용한 논문 이 논문을 다운로드한 분들이 이용한 다른 논문입니다.

      페이지 저장