This paper studies the characteristics of ECG signal. In order to complete ECG signal acquisition, the electrocardiograph front-end circuit includes instrumentation amplifier circuit, high-pass filter circuit, 50 Hz notch circuits, four-pass amplification circuit and low-pass filter circuit. To drive subsequent circuit accurately recorded ECG signal, through the Multisim simulation verifies the feasibility of the circuit, which are verified availability through the experimental test.
목차
Abstract 1. Introduction 2. Design for Front-End Electrocardiograph 2.1. Design of Instrumentation Amplifier 2.2. Design of High-Pass Filter circuit 2.3. 50 Hz Notch Circuit Design 2.4. Main Amplifier Circuit 2.5. Low-Pass Filter 2.6. System Circuit Design 3. System Circuit Simulation 4. System Alignment Test 5. Conclusion References
키워드
ECG signal acquisitionamplifyfilterMultisim simulation