일반인 및 의료종사자들의 자동제세동기를 이용한 제세동에서 전극부착 위치에 관한 연구 : 마네킨을 이용한 연구
Automated external defibrillator electrode pad placement by the general public and health-care providers : A manikin study
Purpose: This study aimed to assess defibrillator pad positioning by the general public and healthcare providers during basic life support (BLS) renewal education (RE). Methods: A total 130 subjects performed defibrillation using an automated external defibrillator (AED) equipped with electrode pads developed by the researchers and male resuscitation manikins, in order to assess electrode pad placement (EPP) before BLS RE. Subjects included 54 Healthcare providers (HP) and 76 members of the general public (GP). Results: The apex EPP was positioned in the risk area (RA), more than 5 cm distant from the area recommended, by 63.2% of the GP (n = 48). The apex EPP was positioned in the RA by 44.4% of the HP (n = 23). The distance from the apex (GP 5.27±2.21 cm, HP 4.18±1.71 cm) was statistically significant, using Student’s t-test. Conclusion: For both GP and HP, apex EPP was significantly in error. Thus, to prevent defibrillation failures, EPP education and practice for GP and HP in all BLS education programs should be improved, and incorrect Korean cardiopulmonary resuscitation guidelines and related literature should be revised and corrected.
목차
Abstract Ⅰ. 서론 1. 연구의 필요성 Ⅱ. 연구방법 1. 연구의 대상 및 기간 2. 연구의 도구 3. 연구의 방법 4. 분석방법 Ⅲ. 연구결과 1. 대상자의 일반적 특성 2. 일반인 및 의료종사자간의 패드부착위치 및 이격거리 비교 Ⅳ. 고찰 Ⅴ. 결론 및 제언 References
한국응급구조학회 [The Korean Society of Emergency Medical Technology]
설립연도
1997
분야
의약학>응급의학
소개
복합 구조적인 산업사회의 발전에 따라 안전사고와 응급상황에서의 구조구급요청자들이 증가하고 있으나 우리나라의 응급의료체계는 아직 그 수준을 따라가지 못하고 있는 실정이다. 이에 우리나라의 응급의료체계를 확립하고 구조구급에 관한 기초학문의 발전에 기여하고자 학회를 창립한다.
간행물
간행물명
한국응급구조학회지(구 한국응급구조학회논문지) [The Korean Journal of Emergency Medical Services ]