백색광과 청색광 스캐너를 이용한 지대치 인상체 스캐닝의 반복재현성 비교
Comparison of reproducibility of prepared tooth impression scanning utilized with white and blue light scanners
Purpose: The purpose of this study compared of reproducibility of prepared tooth impression scanning utilized with white and blue light scanners. Methods: To evaluate reproducibility with white and blue light scanners, the impression of premolar were rotated by 10°~20° and scanned. These data were compared with the first 3-D data (STL file), and the error sizes were measured (n=5). Independent t test was used to evaluation the reproducibility of impression of premolar with white versus blue light scanners through discrepancies of mean, RMS (α=0.05). Results: Discrepancies of mean with regard to reproducibility were 11.2㎛, 5.8㎛, respectively, with white and blue light scanners (p<0.047). And discrepancies of RMS with regard to reproducibility were 33.4㎛, 18.8㎛, respectively, with white and blue light scanners (p<0.045). Conclusion: Our results indicate a good reproducibility of prepared tooth impression digitized with blue light scanner more than that with white light scanner.
목차
Abstract Ⅰ. 서론 Ⅱ. 연구 방법 1. 지대치 모형의 준비 및 인상체 제작 2. 반복재현성 확인을 위한 지대치 인상체 스캐닝 3. 중첩 소프트웨어를 이용한 분석 4. 통계처리 Ⅲ. 결과 Ⅳ. 고찰 Ⅴ. 결론 REFERENCES