Purpose: The purpose of this study was to compare the accuracy of the scan data acquired by the extra-oral and intra-oral scanner. Methods: The maxillary right first molar was made of polymethyl methacrylate (PMMA)specimen. This PMMA specimen was scanned with a engineering scanner and intra-oral scanner. Meanwhile, extra-oral scanner scanned stone die duplicated from PMMA master die. Trueness and precision of scan datas was measuredby3-dimensinal inspection. Independent t-test was conduct to analysis the significant difference.(a=0.05) Results: In the trueness analysis, mean of discrepancies were13.82um for intra oral scanner and16.84 um for extra-oral scanner. In the precision analysis, mean of discrepancies were 11.72 for inta-oral scanner and 9.2 for extra-oral scanner. Both trueness and precision showed a statistically significant difference (Table 1, p<0.05). Conclusion: Intra-oral scanner can show higher trueness than extra-oral scanner, it has lower precision.
목차
Abstract Ⅰ. 서론 Ⅱ. 연구 방법 1. 지대치 제작 2. 스캔 데이터 취득 3. 3차원 분석을 이용한 스캔 데이터의 정확도 분석 4. 통계 분석 Ⅲ. 결과 Ⅳ. 고찰 Ⅴ. 결론 REFERENCES
키워드
Intra-oral scannerExtra-oral scannerAccuracy
저자
이재준 [ Jae-Jun Lee | 고려대학교 대학원 보건과학과 치의기공전공 ]
정일도 [ Il-Do Jeong | 고려대학교 대학원 보건과학과 치의기공전공 ]
김총명 [ Chong-Myung Kim | 고려대학교 대학원 보건과학과 치의기공전공 ]
박진영 [ Jin-Young Park | 고려대학교 대학원 보건과학과 치의기공전공 ]
김지환 [ Ji-Hwan Kim | 고려대학교 대학원 보건과학과 치의기공전공 ]
김웅철 [ Woong-Chul Kim | 고려대학교 대학원 보건과학과 치의기공전공 ]
교신저자