Jong-Dae Kim, Chan-Young Park, Ji-Min Kim, Yu-Seop Kim, Hye-Jeong Song
언어
영어(ENG)
URL
https://www.earticle.net/Article/A207664
※ 원문제공기관과의 협약기간이 종료되어 열람이 제한될 수 있습니다.
원문정보
초록
영어
This paper presents a temperature measurement system without the input buffer. A voltage follower or a buffer is required to minimize the influence of the ADC input impedance in the thermistor temperature measurement system. In such system, a high performance operational amplifier is usually required to prevent the noise and distortion due to the additional circuit. Negative power is also necessary to prevent the distortion of the low voltage input signal. This paper introduces a buffer-less system with a calibration method to resolve such problems. The low temperature coefficient precision resistors were employed to calibrate the input resistance and the leakage current of ADC. The experimental results show that the calibrated error decreased from 190mK to 40mK. As a result, it was shown that the ADC input impedance is time-independent.
보안공학연구지원센터(IJCA) [Science & Engineering Research Support Center, Republic of Korea(IJCA)]