Quartz Crystal Microbalance 시스템을 이용한 저밀도 지질단백질측정용 면역센서의 개발
Development of Quartz Crystal Microbalance-Based Immunosensor for the Determination of Low-Density Lipoprotein
Immunosensor for the determination of LDL(Low-Density Lipoprotein), a good indicator for the diagnosis of atherosclerosis and hypercholesterolemia, was developed by using quartz crystal microbalance(QCM). The immunosensor consists of flow-through cell, oscillating circuit, oscilloscope, and frequency counter. FIA(Flow Injection Analysis) was applied to the QCM system for the measurement of LDL in liquid phase. Antibody showing binding affinity against LDL was immobilized on the gold electrode of a quartz crystal by covalent coupling via polyethylenimine / glutaredehyde. LDL was injected and bound to the antibody immobilized on the QCM immunosensor. The response of the immunosensor (F0 - F1) was found to be proportional to the LDL concentration from 200ug /ml to 800ug /ml. Operational conditions for the operation of immunosensor were also investigated in terms of sensitivity and non-specific binding.
목차
서론 재료 및 실험 재료 QCM 시스템 Flow-through Cell 항체의 고정화 방법 측정방법 결과 및 고찰 응답 특성 곡선 고정화방법들 간의 민감도(sensitivity) 비교 비특미적 결합에 의한 효과 비교 항원-항체 결합에 의한 효과 비교 항체의 농도에 따른 고정화 효율의 최적화 ELISA 에 의한 QCM 시스템내의 항원-항체 결합 역가 확인 요약 감사 참고문헌
한국생물공학회 [The Korean Society for Biotechnology and Bioengineering]
설립연도
1984
분야
공학>생물공학
소개
이 법인은 생물 공학의 발전과 보급에 이바지하고, 회원 상호 간의 연구 협력과 친목을 도모함을 목적으로 한다
1. 생물공학 분야의 발전을 위한 연구 협력
2. 생물공학의 실용화를 촉진시키기 위한 산학 협동
3. 학술연구 발표회, 강연회, 연수회 등 학술활동의 개최
4. 국,영문 학술지,소식지,학술회의 Proceedings 및 학술도서의 발간
5. 생물공학 발전을 위한 정책 건의
6. 기타 국제 교류 등 본 학회의 목적 달성을 위한 제반 활동